Service LibraryPlanning Tools
Contact Russ
Call 772-539-1971Email Russ Jones

Ocala Shop Calibration

Bring the instrument to a controlled, familiar workspace.

MJM is developing an in-shop pressure and temperature calibration capability for suitable portable instruments. We define the requested points, tolerance basis, reference standards, readings, result, and turnover record before accepting the work.

In-house capability

Pressure

Portable pressure gauges, transmitters, indicators, and related devices may be accepted when the requested range, medium, connection, accuracy, and reference capability fit the controlled shop matrix.

In-house capability

Temperature

Portable temperature indicators, probes, switches, and related devices may be accepted when the range, sensor type, immersion needs, accuracy, and comparison method fit the controlled shop matrix.

Not currently advertised

Field and specialty work

Installed loops, field commissioning, mass, electrical, flow, analytical, high-accuracy, or accredited-scope needs require separate review and may be referred to an appropriate provider.

MJM QAAI oversight

A calibration record should explain what was measured and how the result is supported.

01 Identify Instrument, serial or asset ID, range, condition, connection, requested points, tolerance, and intended use.
02 Control Method, environmental conditions, reference standard, calibration status, points, repetitions, and tolerance basis.
03 Review As-found readings, measurement result, uncertainty basis when applicable, pass/fail rule, adjustment authorization, and as-left readings.
04 Package Certificate or service report, supporting data, deviations, limitations, reviewer, and customer turnover.

TECHNICAL LEARNING

Watch a controlled calibration workflow.

Select a measurement type to see a manufacturer demonstration. The equipment shown is representative; MJM confirms the applicable method, range, accuracy, reference, and record before accepting work.

Pressure calibrationFluke demonstrates connection, zeroing, test-point configuration, automatic testing, and result documentation with a pressure calibrator.Watch Fluke procedure
Measurement principle

Traceability belongs to the result

NIST explains traceability as a property of a measurement result supported by a documented, unbroken calibration chain, with each step contributing to uncertainty. A calibrated instrument alone does not establish every later result as traceable.

NIST traceability policy
Reporting principle

Conditions and uncertainty matter

The method, operator, environment, reference equipment, and instrument under test can all contribute to a result. The record must state what MJM actually measured and the limitations that apply.

NIST calibration policies
Accreditation boundary

ISO/IEC 17025

ISO/IEC 17025 is used to assess testing and calibration laboratory competence, impartiality, and consistent operation. MJM does not claim accredited status or an accredited result unless the accepted scope explicitly establishes it.

ISO/IEC 17025 overview
Capability before promise

Available ranges, accuracy or uncertainty needs, reference standards, environmental limits, methods, adjustment authority, accreditation requirements, and reporting expectations are confirmed before an instrument is accepted. No NIST endorsement is implied.

Have a pressure or temperature calibration need?

Send the instrument type, manufacturer/model, range, connection, requested points, tolerance or accuracy requirement, required date, and whether accredited results are required.

Start a shop review